K number
K220958
Device name
Real Intelligence Cori
Country / Region
United States(US)
Product code
OLO
Device class
Class 2
Regulation number
21 CFR 882.4560
Date received
2022-04-01
Decision date
2022-07-27
Submission type
Traditional
Technology type
Software Navigation(AI inferred)
Clinical applications
Knee SurgeryOther
Primary application
Knee Surgery
Classification source
AI classified

Indications for Use

REAL INTELLIGENCE CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include:
• unicondylar knee replacement (UKR),
• total knee arthroplasty (TKA),
• revision knee arthroplasty, and
• total hip arthroplasty (THA).
CORI is intended to be used by trained medical professionals in a hospital or clinical setting equivalent to an orthopedic surgery suite.

Intended Use

REAL INTELLIGENCE CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), total knee arthroplasty (TKA), revision knee arthroplasty, and total hip arthroplasty (THA).

Device Description

The subject of this Traditional 510(k) is REAL INTELLIGENCE CORI (CORI), a robotic-assisted orthopedic surgical navigation and burring system. CORI uses established technologies of navigation via a passive infrared tracking camera. Based on intraoperatively-defined bone landmarks and known geometry of the surgical implant, the system aids the surgeon in establishing a bone surface model for the target surgery and planning the surgical implant location. For knee applications, CORI then aids the surgeon in executing the surgical plan by controlling the cutting engagement of the surgical bur. CORI knee application software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved with two modes: Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the robotic system retracts the bur inside the guard, disabling cutting. Speed control regulates the signal going to the tool control unit itself and limits the speed of the drill if the target surface is approached. Alternatively, the surgeon can disable both controls and operate the robotic drill as a standard navigated surgical drill.

Summary parsed at:2026-06-24 04:34:35 UTC

Predicate devices

23

K220958 references 23 predicate/reference devices

Cited as predicate

0

No later 510(k)s cite this device yet

Other submissions will appear here when they list K220958 as a predicate